مقاله
|
Abstract
|
|
|
Title:
|
wavelight topolyzer vario VS Javal keratometer; Comparison of Keratometric Values
|
Author(s):
|
Ali Ayatollahi, Jahangir Ayatollahi
|
Presentation Type:
|
Oral
|
Subject:
|
Cornea and Anterior Segment
|
Others:
|
|
Presenting Author:
|
|
Name:
|
Ali Ayatollahi
|
Affiliation :(optional)
|
Novindidegan Eye Center
|
E mail:
|
a.ayatollahi@yahoo.com
|
Phone:
|
02122144882
|
Mobile:
|
09134502474
|
|
|
Purpose:
|
The purpose of this study was to compare keratometry measurements from the wavelight topolyzer vario and the Javal Manual keratometer.
|
Methods:
|
Keratometry measurements in the steep (Ks) and flat (Kf) meridians were taken with the Javal Manual keratometer (Javal Schiotz type; Haag-Streit AG, Koeniz,Switzerland) and wavelight topolyzer vario ( Wavelight, Germany)in myopic virgin eyes. The 95% limits of agreement (LOAs) were reported to evaluate the agreement between devices.
|
Results:
|
The mean K readings of wavelight topolyzer vario and Javal Manual keratometer were 43.89/43.71 diopter (D), 44.01/43.92 D, respectively. Correlation coefficients for the CASIA AS-OCT and Javal Manual keratometer in measuring flat K and steep K were r=0.918 and r=0.921, respectively (p<0.001).
|
Conclusion:
|
The wavelight topolyzer vario provided reliable K
measurements in clinical practice in comparison with Javal Manual keratometry and seem to be interchangeable.
|
Attachment:
|
|
|
|